- Home
- Products
- Process Measurements
- Teledyne Analytical Instruments
- Portable Oxygen Analyzer 3110
Portable Oxygen Analyzer 3110
Portable Oxygen Analyzer
- General Details
- Options
- Standard Features
- Technical data
- Documents
The Model 3110 represents the new generation of portable trace and percent oxygen analysis capability. The 3110 combines a rugged, portable analyzer design with the high reliability of Teledyne trace and percent level sensors. This ensures highly accurate ppm/percent oxygen readings in a variety of background gases — including hydrocarbons. The batteries supply at least 100 hours of continuous power to the analyzer, with a low-battery indicator LED. The analyzer includes quick-disconnect fittings to take measurements without hassle and can be equipped with an external sample system. The microprocessor-based electronics provide accurate, high resolution readings and easy-to-use features.
• Real-time data-logging capabilities with digital output that can be downloaded directly to a PC
• Stainless steel quick disconnect gas fittings
• Sample system consisting of coalescing filter, needle valve, flow meter, fittings and tubing suitable for trace/percent O2 analysis
• Insta-trace for fast recovery after installation
• NG option: H2S and sulfur resistant sensor
- Meets technical requirements of ASTM D7607-11 (3110XL only)
- User-settable ranges
- High accuracy and sensitivity; fast response
- Unaffected by hydrocarbons and other oxidizable gases
- Ideal for measuring oxygen in hydrogen, nitrogen, argon, helium, acetylene, ethylene, butadiene, and many other gases
- Long-life, maintenance-free, Micro-fuel Cell oxygen sensor
- No support gases required
- Rugged, lightweight, and reliable
- Carrying case included
Ranges | : | 3110: 0-10 ppm to 0-25% O2 |
3110XL: 0-2 ppm to 0-25% O2 | ||
3110P: 0-1% to 0-25% O2 | ||
Response Time | : | Trace ranges (with B-2C sensor): < 60 seconds or less |
Percent ranges (with B-1 sensor): < 7 seconds | ||
Operating Temperature | : | 32°F to 122°F (0°C to 50°C) |
Reproducibility | : | ±1% at constant temperature |
Sensor Type | : | Micro-fuel Cell |
System Power | : | NiCad batteries, AC adapter included for battery charging 110/220 VAC 50/60 Hz, 0.25A |
Weight | : | 6 lbs. (2.71 kg) |
Approval | : | Factory Mutual (FM/FMc), intrinsically safe, Class I Div 1 Groups A-D T6 ATEX, intrinsically safe, Ex ia IIC T3 CE marked |
Battery | : | Ni-Cad, >100 hours per charge, 5-year expected life |