- Home
- Products
- Water Measurements
- Thermo Scientific
- Wastewater Analysis
- Orion™ AquaMate 8000 UV-Vis Spectrophotometer
Orion™ AquaMate 8000 UV-Vis Spectrophotometer
- General Details
- Specification
- Features
- Documents
General Details
Equip your lab with the accurate, easy to use Thermo Scientific™ Orion™ AquaMate 8000 UV-Vis Spectrophotometer to ensure you have the ideal instrument for water analysis and beyond. This spectrophotometer offers 260 preprogrammed methods for ion concentration from absorbance plus wavelength scanning function and standard curve function for creating custom method using so virtually any brand of colorimetric reagent can be used. Utilize 190-1100 nm selectable wavelengths, 1.8 nm spectral bandwidth and vial flexibility with the included 3-position carousel for holding round and rectangular vials .
Specification
Accuracy (Photometric) | ±0.005A at 1.0A; |
Compatibility | Compatible with ink jet and laser printers with HP PCL 5 or greater |
Connections | USB type A port for USB stick (front panel), USB type B port for computer (rear panel), USB type A port for printer (rear panel) |
Description | AquaMate 8000 UV-Vis Spectrophotometer |
Detector Type | Dual silicon photodiodes |
Display | Graphical with LCD backlight—9.7 x 7.1cm (3.8 x 2.8 in.) |
Drift | <0.0005A/hr |
Electrical Requirements | 100 to 240V 50/60Hz |
Height (English) | 9.8 in. |
Height (Metric) | 25cm |
Hertz | 50/60Hz |
Includes | Methods on USB memory stick; 6-position cell holder; 3-position cell holder; 1-position cell holder; User documentation; Dust cover; USB cable; 110V and 220V power cords |
Item Description | AquaMate 8000 UV-Vis Spectrophotometer |
Keypad | Sealed membrane with tactile response keys |
Lamp | Xenon flash lamp (5 years) |
Min. Data Interval | 0.2; 0.5; 1.0; 2.0; 3.0; 5.0nm |
Noise | |
Optical Design | Dual beam—internal reference detector |
Photometric Linearity | Up to 3.5A at 260nm |
Range (Photometric) | -0.5 to 5.0 A ; -1.5 to 125 %T; ±9999 C |
Spectral Bandwidth | 1.8nm |
Stray Light | |
Voltage | 100 to 240V |
Wavelength Accuracy | ±1.0nm |
Wavelength Range | 190 to 1,100nm |
Wavelength Repeatability | ±0.5nm |
Wavelength Scan Speed | 10 to 4,200nm/min |
Wavelength Slew Speed | 11,000nm/min |
Weight (English) | 19 lb. |
Weight (Metric) | 8.6kg |
Width (English) | 11.8 in. |
Width (Metric) | 30cm |
Features
Pre-programmed Methods:
- More than 260 pre-programmed methods offer easy and convenient measurements using Orion™ AQUAfast™, Merck™ or CHEMetrics™ chemistries
- Pre-programmed methods can also be adapted for new chemistries or operators can create their own custom methods and calibration curves
- Instruments allow a one-point adjustment on any preprogrammed method to correct for variations in batch reagent chemistries
Versatility:
- Wide variety of vials sizes can be used, including 10mm square vials, 20 and 50mm rectangular vials and an adjustable round vial holder that accommodates 13, 16, and 24mm round vials
- Three USB connections provide the options to store methods and data to a USB memory stick, transfer data to a computer, and print hard copy data to an external printer
- System is compatible with ink jet and laser printers with HP PCL 5 or greater
Verification:
- Performance verification tests ensure wavelength accuracy and instrument functionality
- Each verification report gives the time, date and instrument serial number in accordance with GLP and GMP
- Built-in wavelength accuracy test is compatible with either the internal lamp or external calibrated standards
- Built-in test methods for stray light, noise and resolution are available to validate instrument performance further
Includes:
- Methods on USB memory stick
- Three-position cell holder
- User documentation
- Dust cover
- USB cable
- 110V, 220V and 240V power cords