- Home
- Products
- Process Measurements
- H2scan Corp
- Hydrogen measurements
- HY-OPTIMA™ 1720 Intrinsically Safe In-line Process Hydrogen Analyzer (Oxygen & Nitrogen)
HY-OPTIMA™ 1720 Intrinsically Safe In-line Process Hydrogen Analyzer (Oxygen & Nitrogen)
- General Details
- Documents
H2scan’s HY-OPTIMA™ 1720 in-line, real-time hydrogen specific process analyzers are designed for ease of use, interface flexibility and process control. The HY-OPTIMA™ 1720 is an intrinsically safe, solid-state sensor that is configured to operate in air background gas streams with up to 95% relative humidity and temperatures up to 60ºC. It is ideal for the detection of hydrogen within processes where real-time monitoring can increase plant safety, improve process diagnostics and enhance maintenance management.
Performance
Hydrogen Sensitivity Range:
0.4% to 5% hydrogen by volume at 1 ATM
Typical Response Time: T90 of 60 sec. maximum
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy: ± (0.03 x indication + 0.2) percent
Hydrogen-by-volume
Interface Options
Power Barrier Input Voltage: 24 VDC nominal
Power Barrier Input Voltage Range: 20 VDC to 28 VDC
Intrinsically Safe Sensor Input Voltage: 10 VDC
Sensor Input Voltage Range: 5 VDC to 28 VDC
Input Power: 10 W
Analog Output: 4 mA to 20 mA
Serial Communication Options: RS422
Certifications
CE 0359, Ex
II 1 G, Ex ia II H2, d+d IIB T3
II 2 G, Ex ib II H2, ib d IIB T3
ITS07ATEX25634X
Operating Conditions
Process Gas Temperature Range: -20ºC to 60ºC
Process Gas Humidity: 0% to 100%
Gas Flow Rate: 0.1 slpm to 10 slpm
Gas Pressure: 0 to 2 bar gauge, 0 to 30 psi gauge
Operating Temperature Range: -20ºC to 40ºC
Operating Humidity: 0% to 95% RH (non-condensing)
Storage Temperature Range: -40ºC to 50ºC
Calibration Background Gas: Oxygen & Nitrogen
Included Accessories
Intrinsically Safe Power Barrier
Armored Power / Analog Cable: 4 m
Calibration Cup
Optional Accessories
Single Channel Intrinsically Safe Analog Output Barrier (with or without relays)
Single Channel Intrinsically Safe RS422 Serial Interface Barrier
Serial Interface Cable: 4 m (custom lengths available)
Armored Power/Analog Cable (custom lengths available)
Adapter Fittings
½ in. MNPT thread
½ in. FNPT thread
-8 SAE/MS thread size
… and many other industry standards.
Dimensions
Total Length: 10.3 in.
Length of Enclosure: 6.8 in.
Length of Tube: 2.9 in.
Width: 3.4 in.
Please Us for the latest datasheet/product manual