9110T / 9110TH Chemiluminescence NOx Analyzers

General Details

The Model 9110T uses the proven chemiluminescence detection principle, coupled with state-of-the-art microprocessor technology to provide the sensitivity, stability and ease of use needed for ambient or dilution CEM requirements for monitoring levels of oxides of nitrogen, thermal DeNOx, stack testing and NOx testing in beverage-grade carbon dioxide. 

The analyzer platform uses multi-tasking software which allows complete control of all functions while providing online indication of important operating parameters. Measurements are automatically compensated for temperature and pressure changes. In addition, stability is enhanced by an auto-zero circuit that continuously provides a true zero reference. The result is sensitive, accurate, and dependable performance under the harshest operating conditions. 

Selectable measurement modes (NO only, NOx only, NO/NOx switching), auto ranging, and remote range control customizes the analyzer for the end user. The modular design offers top-mounted access to all components while hinged front and rear panels simplify module replacement. The unique modular sealed ozone generator provides improved voltage coefficient, safety and reliability. A standard permeation dryer on the ozone generator means there is no desiccant to replace. In addition, the excess ozone is removed by catalytic reaction, eliminating the need for frequent charcoal scrubber replacement. 

ADVANCED PERFORMANCE
The T-Series platform offers an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. It includes a built-in data acquisition capability using the analyzer’s own internal memory. This allows the logging of multiple parameters including averaged or instantaneous concentration values, calibration data and operating parameters such as flow, pressure and lamp intensity. 

Stored data are easily retrieved through the front panel or via RS232, Ethernet, or USB ports either locally or by remote connection using the included APIcomTM software. These innovative features allow operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends. The 9110T combines lighter weight, rugged construction, ease of use, powerful diagnostics, modular design and outstanding performance to meet air monitoring requirements.

Options

• Real-time data-logging capabilities with digital output that can be downloaded directly to a PC
• Stainless steel quick disconnect gas fittings
• Sample system consisting of coalescing filter, needle valve, flow meter, fittings and tubing suitable for trace/percent O2 analysis
• Insta-trace for fast recovery after installation
• NG option: H2S and sulfur resistant sensor

Standard Features

• Ranges: 0-50 ppb to 0-20 ppm (9110T) or 0-5 ppm, 0-5,000 ppm (9110TH), userselectable
• Large, vivid, and durable color graphics display with touch screen interface
• Independent rangesfor NO, NO2 , and NOx
• Front panel USB connectionsfor peripheral devices and firmware upgrades
• Integral permeation dryer and catalytic ozone scrubber
• Temperature and pressure compensated
• Optional oxygenator for O2 -free samples – optimal for CO2 purity analysis

Specification 9110T
General 9110T
Ranges Min: 0-50 ppb Fullscale Max: 0-20,000 ppb Fullscale (selectable, independent NO, NO2 , NOx ranges and auto ranging supported)
Measurement Units ppb, ppm, μg/m3, mg/m3 (selectable)
Zero Noise < 0.2 ppb (RMS)
Span Noise < 0.5% of reading (RMS) above 50 ppb
Lower Detectable Limit 0.4 ppb
Zero Drift < 0.5 ppb/24 hours
Span Drift < 0.5% of fullscale/24 hours
Lag Time 20 seconds
Rise and Fall Time < 60 secondsto 95%
Linearity 1% of fullscale
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 500 cm3/min ±10%
ELECTRICAL
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
COMMUNICATIONS
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device port
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs(3 x 4-20mA current outputs)
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs(0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
PHYSICAL
Operating Temperature Ranges 5 – 40ºC
Dimensions (HxWxD) 7”x 17”x 23.5”(178 x 432 x 597 mm)
Weight Analyzer: 40 lbs(18 kg)
External pump: 15 lbs(7 kg)
Specification 9110TH
General 9110TH
Ranges Min: 0-50 ppb Fullscale Max: 0-5,000 ppb Fullscale (selectable, independent NO, NO2 , NOx ranges and auto ranging supported)
Measurement Units ppm, mg/m3 (selectable)
Zero Noise < 20 ppb (RMS)
Span Noise < 0.2% of reading (RMS) above 20 ppm
Lower Detectable Limit 40 ppb
Zero Drift < 20 ppb/24 hours
Span Drift < 0.5% of fullscale/24 hours
Lag Time 20 seconds
Rise and Fall Time < 60 secondsto 95% (in switching mode)
Linearity 1% of fullscale
Precision 0.5% of reading above 50 ppb
Sample Flow Rate 290 cm3/min ±10%
ELECTRICAL
Power Requirements 100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges 10V, 5V, 1V, 0.1V (selectable)
Recorder Offset ±10%
COMMUNICATIONS
Included I/O 1 x Ethernet: 10/100Base-T
2 x RS232 (300-115,200 baud)
2 x USB device port
8 x opto-isolated digital outputs
6 x opto-isolated digital inputs
4 x analog outputs(3 x 4-20mA current outputs)
Optional I/O 1 x USB com port
1 x RS485
8 x analog inputs(0-10V, 12-bit)
4 x digital alarm outputs
Multidrop RS232
PHYSICAL
Operating Temperature Ranges 5 – 40ºC
Dimensions (HxWxD) 7”x 17”x 23.5”(178 x 432 x 597 mm)
Weight Analyzer: 40 lbs(18 kg)
External pump: 15 lbs(7 kg)