7500Z / 7600Z Non-Dispersive Infrared (NDIR) Multi-Gas Analyzers
Teledyne Analytical Instruments
- General Details
- Product features:
- Specification 7500ZA/7500ZB
- Specification 7600ZA/7600ZB
- Specification 7600ZA-XL/7600ZB-XL
- Documents
Teledyne’s 7500Z and 7600Z Series Infrared (IR) Analyzers are capable of detecting up to four chosen, individual IR absorbing components (i.e. CO, CO2, CH4, SO2, NO or NOX with external converter) on a continuous basis. Conveniently packaged in either a 19” rack mount (Model 7500ZA / 7600ZA) or NEMA-4 wall mount enclosure (Model 7500ZB / 7600ZB), the analyzer can also be supplied with an oxygen sensor, providing the operator with a space saving, five in one, cost-effective design. The NEMA-4 enclosure can be purged to meet Class I Div 1 (Zone 1) or Class I Div 2 (Zone 2) hazardous area installation requirements.
A high-sensitivity mass flow type twin detector is utilized for infrared measurements. By using a single beam single path design in conjunction with a serial dual-layer transmission detector, the Series 7500Z delivers long term, drift-free performance. The Series 7600Z achieves greater performance through an internal sample-switching system which serves to cancel zero drift.
The oxygen concentration in the sample gas can be detected by either a built-in paramagnetic sensor or electrochemical sensor.
The concentration of the desired gases is displayed on a large, easy-to-read back-lit LCD. The user interface is very intuitive and the menu / mode selection buttons, which are readily accessible, provide the operator with dynamic control and extensive diagnostic capabilities.
• Up to five (5)simultaneous, continuous measurements of CO, CO2 , NO / NOX , SO2 , CH4 and O2
• Rangesfrom 0-10 ppm to 0-100% (gas-dependent)
• Luminescent LCD displayshowing all measurements and settings
• Low-profile 3U design to minimize rack space
• Oxygen measurement with paramagnetic or electrochemical sensor (optional)
SPECIFICATIONS | 7500ZA/7500ZB |
REPEATABILITY | ±0.5% of fullscale |
LINEARITY | 1% of fullscale prior to atmospheric pressure correction (option) |
ZERO DRIFT | ±2.0% of fullscale/week In the case of Auto Zero calibration use for 500ppm or lessrange |
SPAN DRIFT | ±2% of fullscale/week |
RESPONSE TIME (T90) |
1 to 15 sec electrical response. Within 10-30 secondsincluding replacement time ofsampling gas. Gasreplacement time depends on the number of measuring components, and measuring range |
OPERATING TEMPERATURE | 23 to 113°F (-5 to 45°C) |
SAMPLE PRESSURE | Lessthan 1.5 psig, vent to atmospheric pressure |
SAMPLE FLOW |
1.0 SCFH (0.5 SLPM) |
UTILITY GASES | Not required |
POWER | 100 – 240 VAC, 50 / 60 Hz, 100 VA max |
RACK-MOUNT DIMENSIONS & WEIGHT |
19.00” L x 5.22” H x 15.04” D (483 x 133 x 382 mm) Approx. 25 pounds |
WALL-MOUNT DIMENSIONS & WEIGHT |
24.00”L x 24.00”H x 12.00”D (610 x 610 x 305 mm), Approx. 60 pounds |
SPECIFICATIONS | 7600ZA/7600ZB |
REPEATABILITY | ±0.5% of fullscale |
LINEARITY | 1% of fullscale prior to atmospheric pressure correction (option) |
ZERO DRIFT | ±0.5% of fullscale/week (measurable component of NDIR) ±2.0% of fullscale/week (O2 analyzer) |
SPAN DRIFT | ±2.0% of fullscale/week |
RESPONSE TIME (T90) |
30 seconds or better Response interval may be changed depending on timing of the switching gas bysample switching operation. (Td=5-20 seconds) |
OPERATING TEMPERATURE | 23 to 113°F (-5 to 45°C) |
SAMPLE PRESSURE | Lessthan 1.5 psig, vent to atmospheric pressure |
SAMPLE FLOW |
1.0 SCFH (0.5 SLPM) |
UTILITY GASES | Not required |
POWER | 100 – 240 VAC, 50 / 60 Hz, 100 VA max |
RACK-MOUNT DIMENSIONS & WEIGHT |
19.00” L x 5.22” H x 15.04” D (483 x 133 x 382 mm) Approx. 29 pounds |
WALL-MOUNT DIMENSIONS & WEIGHT |
24.00”L x 24.00”H x 12.00”D (610 x 610 x 305 mm), Approx. 60 pounds |
SPECIFICATIONS | 7600ZA-XL/7600ZB-XL |
REPEATABILITY | ±0.5% of fullscale |
LINEARITY | 1% of fullscale prior to atmospheric pressure correction (option) |
ZERO DRIFT | ±0.5% of fullscale/week (measurable component of NDIR) ±2.0% of fullscale/week (O2 analyzer) |
SPAN DRIFT | ±2.0% of fullscale/week |
RESPONSE TIME (T90) |
30 seconds or better Response interval may be changed depending on timing of the switching gas bysample switching operation. (Td=5-20 seconds) |
OPERATING TEMPERATURE | 23 to 113°F (-5 to 45°C) |
SAMPLE PRESSURE | Lessthan 1.5 psig, vent to atmospheric pressure |
SAMPLE FLOW |
1.0 SCFH (0.5 SLPM) |
UTILITY GASES | Not required |
POWER | 100 – 240 VAC, 50 / 60 Hz, 100 VA max |
RACK-MOUNT DIMENSIONS & WEIGHT |
19.00” L x 5.22” H x 15.04” D (483 x 133 x 382 mm) Approx. 25 pounds |
WALL-MOUNT DIMENSIONS & WEIGHT |
24.00”L x 24.00”H x 12.00”D (610 x 610 x 305 mm), Approx. 60 pounds |