- Home
- Products
- Process Measurements
- Teledyne Analytical Instruments
- 4060 Gas Chromatographs
4060 Gas Chromatographs
Teledyne Analytical Instruments
- General Details
- Product features:
- Common Detectable Gases 4060/FID | 4060/TC
- Applications
- Specification 4060/FID | 4060/TC
- Documents
The 4060 Series of Gas Chromatograph (GC) analyzers combine Teledyne’s field proven detector technologies with gas chromatography column separation capability. This enables reliable and repeatable of a single or multiple compounds in a stream containing interfering components.
Flame-ionization detector (FID) or thermal conductivity detector (TCD) options enable part-per-billion (ppb), part-per-million (ppm) or percent levels of detection depending on application. The analyzer includes an integral temperature-controlled sample system of 316SS and Teflon construction to provide a suitable sample to the detector. Contact TAI for other available materials.
- Field-proven microprocessor-based electronics with menu-driven functionality
- Easy-to-view 5-digit LED display for readings
- 2 x 20 vacuum fluorescent display for menu options, settings and status
- 0-1 VDC and 4-20mA (isolated) analog outputs and gas identification outputs
- RS-232 duplex digital output
- Two fully-configurable concentration alarms and one system alarm, all with Form-C relay contacts
- Built-in diagnostic capability
- Auto-calibration and remote-calibration ready (integral valves available)
| 4060/FID | 4060/TC |
| Acetaldehyde | Argon (purity and impurity levels) |
| Benzene | Carbon Dioxide |
| Butane | Carbon Monoxide |
| Carbon Dioxide – (measurement achieved with methanizer) | Helium |
| Carbon Monoxide – (measurement achieved with methanizer) | Hydrogen |
| Ethane | Hydrogen Sulfide |
| Ethyl Benzene | Krypton |
| Ethylene | Nitrogen (purity and impurity levels) |
| Hexane | Xenon |
| Methane | |
| Methanol | |
| MTBE | |
| Propane | |
| Propylene | |
| Toluene | |
| VOCs | |
| Xylene | |
- Ambient benzene, BTEX detection
- Beverage-grade carbon dioxide: detection of ppb benzene, acetaldehyde, methanol and VCM in beverage-grade carbon dioxide to ISBT specification
- Methane / non-methane hydrocarbon monitoring
- Nitrogen or Argon purity
- C1-C4 levels in natural gas or syngas production
- LNG impurity detection
- MTBE monitoring in gasoline
- Hydrogen detection in a mixed fuel gas
Chromatography Software – Tracs™
Tracs™ software developed by TAI provides all features and functionality needed to view and analyze chromatography data. When installed on a PC computer, Tracs™ collects digitized data fed from the analyzer with which it can plot chromatograms, perform analytical calculations, log data, and save information for future use.
Features include:
- Bi-directional communication of analyzer’s range, alarm and timing settings which can be remotely uploaded and downloaded
- Acquires real-time raw data for continuous display of chromatograms
- Chromatograms can be superimposed upon one another to verify repeatability
- Elution peaks can be displayed within their own respective timing frames or individually manipulated for optimal peak scrutinization
- Statistical summary including analytical calculation of measurements, distribution of values and deviations
- Data-logging
- Saving and retrieving of digitized data files
| 4060/FID | 4060/TC | |
| Sensing Technology | Flame-Ionization Detector (FID) | Thermal Conductivity Detector (TCD) |
| Gas and Ranges | Hydrocarbon compounds Trace or percent level detection (Contact factory to verify application) |
Most gases Trace or percent level detection (Contact factory to verify application) |
| Sensitivity | 1% of full-scale | 1% of full-scale |
| Accuracy | ± 2% of full-scale (at constant temperature) |
± 2% of full-scale (at constant temperature) |
| Repeatability | 1% of full-scale through all ranges | 1% of full-scale through all ranges |
| Response Time | Application Dependent | Application Dependent |
| Operating Temperature | 40-110°F (4-43°C) | 40-110°F (4-43°C) |
| Zero and Span Noise | Less than 0.5% of full-scale | Less than 0.5% of full-scale |
| Zero and Span Drift | 2% of full-scale per week | 2% of full-scale per week |
| Analog Outputs | 0-1 VDC and 4-20 mADC isolated (one output with Gas ID for 3+ gases) |
0-1 VDC and 4-20 mADC isolated (one output with Gas ID for 3+ gases) |
| Alarms | One system alarm and two concentration alarms. Form-C relays rated @ 3A 250VAC resistive |
One system alarm and two concentration alarms. Form-C relays rated @ 3A 250VAC resistive |
| Supply Voltage | 110 or 220 VAC 50/60 Hz | 110 or 220 VAC 50/60 Hz |
| Maximum Power Consumption | 600 VA | 600 VA |
| Maximum Load Impedance of 4-20 mA Output |
500 ohms | 500 ohms |
| Sample Flow Rate | 1 SCFH (0.5 LPM) standard | 1 SCFH (0.5 LPM) standard |
| Dimensions | 19” W x 8.75” H x 22.5” D (483 x 222 x 572mm) |
19” W x 8.75” H x 16.5” D (483 x 222 x 419mm) |