- Home
- Products
- Flue Gas Monitoring
- E Instruments International
- Industrial Emissions Analyzers
- E8500 Plus Portable Industrial Combustion Gas & Emissions Analyzer
E8500 Plus Portable Industrial Combustion Gas & Emissions Analyzer
- General Details
- Technical Data
- Product Video
- Documents
General Details
The E8500 Plus is the most powerful and advanced portable emissions analyzer on the market, and is a complete, portable tool for EPA compliance–level emissions monitoring of boilers, engines, and other combustion equipment.
- Up to 9 Gas Sensors: O2, CO, NO, NO2, SO2, H2S (Electrochemical), CO2, CxHy Hydrocarbons, High CO (NDIR), VOC (PID)
- Low NOx and “True” NOx Measurements
- Built–in Thermoelectric Chiller with Automatic Condensate Removal
- Software for Real–Time Data Logging, Graphing, and Reporting with unlimited memory
- Blue Tooth Wireless Communications
- CO Dilution Auto–Range for Measurements up to 20,000ppm
- Automatic Long–Term Monitoring & Data Logging
- Enhanced Internal Memory (2000 Tests)
- Sample Conditioning Unit for Low NOx & SO2 Measurements
- Bluetooth Wireless Remote Printer
- Long Life Li-Ion Rechargeable Battery
- High Temperature Probes with Sintered Filter
- Stack Gas Velocity with Pitot Tube
- Temperature & Pressure Measurements
- CO2%, Efficiency, & Excess Air Calculations
- Light Weight, Durable, & Easy to Transport Design with Easy Filter Access
- Complies with USEPA CTM-030 & CTM-034 test methods
Technical Data
E8500 Plus | |
---|---|
Maximum Number of Sensors | 9 |
Internal Memory (Samples Stored) |
2000 |
Draft/Differential Pressure | Standard |
Gas Velocity | Optional |
Built-In Printer (Non-Thermal) |
Optional Remote Wireless Printer |
O2 Sensor (0-25%) |
Standard |
CO Sensor (0-8000ppm) |
Standard |
CO Dilution Auto-Range (automatic) | Standard, up to 20,000 ppm |
CO2 Calculated (0-99.9%) |
Standard |
NO Sensor (0-5000ppm) |
Standard |
NO2 Sensor (0-1000ppm) |
Optional |
SO2 Sensor (0-5000ppm) |
Optional |
Low NOx/SO2 Sensors (0-100.0ppm) |
Optional |
CxHy(HC) Sensor Pellistor type (0-5%) |
|
H2S Sensor (0-500ppm) |
Optional |
VOC Sensor (0-200ppm) |
Optional |
CO2 NDIR Sensor (0-20%) |
Optional |
CxHy(HC) NDIR Sensor (0-3%) |
Optional |
CO NDIR Sensor (0-15%) |
Optional |
Built-In Peltier Chiller | Standard |
Automatic Data Logging | Standard |
Bluetooth Wireless Communications | Standard |
Software Package | Standard, Real-Time |
E8500 Plus |
Product Video
[youtube id=”e0at33zfCEA” align=”left” mode=”normal” autoplay=”no” aspect_ratio=”16:9″ maxwidth=”400″ parameters=”https://www.youtube.com/watch?v=e0at33zfCEA” grow=”yes”]
Documents