HY-OPTIMA™ 730 In-line Process Hydrogen Analyzer (ppm H2S & CO)

General Details

H2scan’s Series 700 in-line, real-time hydrogen specific process analyzers are designed for ease of use, interface flexibility and true process control. The HY-OPTIMA™ 730 is a solid-state sensor configured to operate in background gas streams having carbon monoxide concentrations up to 100 ppm, hydrogen sulfide stream concentrations up to 1000 ppm, up to 95% relative humidity, and temperatures up to 60ºC. The HY-OPTIMA™ 730 is ideal for hydrogen production and petrochemical applications where real-time measurements of hydrogen can increase process plant efficiencies, improve diagnostics and enhance maintenance management translating into higher profitability.

Performance 

Hydrogen Sensitivity Range:

0.5% to 100% hydrogen at 1 ATM on a
volume basis in a single or multi-component
gas background

Typical Response Time: T90 less than 30 sec
Ingress Protection: IP64 capable
Recommended Verification Interval: 90 days
Product Life Expectancy: 10 years
Accuracy(*):
± 0.3% absolute for 0.5 to 10% H2
± 1.0% absolute for 10 to 100% H2
Drift/week:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Repeatability:
± 0.2% absolute for 0.5 to 10% H2
± 0.4% absolute for 10 to 100% H2
Linearity:
± 0.2% absolute for 0.5-10% H2
± 0.4% absolute for 10-100% H2

Interface Options   

Input Voltage Range: 8 VDC to 13 VDC
Input Current: 1.0 A
Analog Outputs:
Output Voltages – OR – Output Currents
• 0 VDC to 5 VDC        • 4 mA to 20 mA
• 1 VDC to 5 VDC        • 0 mA to 20 mA
• 0 VDC to 4 VDC        • User-specific mA range
• 0.5 VDC to 4.5 VDC
• User-specific VDC range
Serial Communication Options: RS232 or RS422
Relay Contacts: Two programmable 60 VDC / 1 A
SPDT relays with both normally open (N.O.)
and normally closed (N.C.) contacts.

Certifications

Operating Conditions

Process Gas Tolerance:
CO tolerance < 100 ppm
H2S tolerance < 1000 ppm
Operating Humidity: 0% to 95% RH
Flow Rate: 0.1 slpm to 10 slpm
Pressure: 0 to 2 bar gauge, 0 to 30 psi gauge
Process Gas Temperature Range: -20ºC to 60ºC
Operating Temperature Range: -20ºC to 40ºC
Storage Temperature Range: -40ºC to 50ºC
Calibration Background Gas: Nitrogen

Adapter Fittings    

½ in. MNPT thread
½ in. FNPT thread
-8 SAE/MS thread size
… and many other industry standards

Dimensions    

Total Length: 9.1 in.
Length of Enclosure: 6.7 in.
Length of Tube: 2.4 in.
Width: 3.4 in.
Depth: 1.4 in.

Documents

Please Contact Us for the latest datasheet/product manual